The outstanding success of the Polilyte Plus in chemical and wastewater applications gave the inspiration for transferring the good features to a whole family of sensors. The expanded portfolio widens the range of applications that can be covered
| ATEX Approval | No | 
|---|---|
| Autoclavable | Yes, Maximum Temperature 130 °C | 
| CIP | Yes | 
| Diameter | 12 mm | 
| Diaphragm | Single Pore | 
| Electrical Connector | VP8 | 
| Electrolyte | Polisolve Plus | 
| Liquid Earth | No | 
| Measurement Principle | Combination electrode; pH potential measured against reference | 
| Measuring Range | pH 0 to 14 | 
| Membrane Shape | Spherical | 
| Number of Diaphragms | 2 | 
| Operating Temperature Range | Analog interface: 0 to 110 °C; Digital interface: 0 to 130 °C | 
| Parameter | pH | 
| Pressure Range | 0 to 16 bar (100 °C), 0 to 10 bar (130 °C) | 
| Process Connection | PG13,5 | 
| Recommended Storage | Hamilton Storage Solution or 3 M KCl | 
| Reference System | Hamilton Everef-L | 
| Sample min Conductivity | 2 µS/cm | 
| Sensitivity | 57 to 59 mV / pH at 25 °C | 
| Sensor Family | Polilyte Plus | 
| Sensor Output | Arc: Modbus, 4 to 20 mA | 
| Serial Number | Yes | 
| Steam Sterilizable | Yes, Maximum Temperature 130 °C | 
| Temperature Sensor | NTC 22 kOhm | 
| Wetted Parts | Glass \nEPDM (Ethylene propylene elastomer) | 
| Zero Point | 0 ± 20 mV | 
| Length | 325 mm | 
| pH Glass | Hamilton type HB glass | 
| pH Glass Resistance | Moderate | 
| Analog Interface 1 | 4 to 20 mA for pH, programmable | 
| Analog Interface 1 and 2 | Galvanically not isolated; pulse width modulation 3.5 kHz | 
| Analog Interface 2 | 4 to 20 mA for Temperature programmable | 
| Baud Rate | 4800, 9600, 19200, 38400, 57600,115200 Bd | 
| Configurable Values | pH: pH, mV; Temperature: °C, K, °F | 
| Digital Interface | RS485 Modbus (Maximum 31 addresses) | 
| Operating Voltage | 7 to 30 VDC Maximum 150 mW |