The outstanding success of the Polilyte Plus in chemical and wastewater applications gave the inspiration for transferring the good features to a whole family of sensors. The expanded portfolio widens the range of applications that can be covered
| ATEX Approval | No |
|---|---|
| Autoclavable | Yes, Maximum Temperature 130 °C |
| CIP | No |
| Diameter | 12 mm |
| Diaphragm | Single Pore |
| Electrical Connector | VP8 |
| Electrolyte | Polisolve Plus |
| IECEx Approval | No |
| Liquid Earth | No |
| Measurement Principle | Combination electrode; pH potential measured against reference |
| Measuring Range | pH 0 to 14 |
| Membrane Shape | Spherical |
| Number of Diaphragms | 2 |
| Operating Temperature Range | Analog interface: 0 to 110 °C; Digital interface: 0 to 130 °C |
| Parameter | pH |
| Pressure Range | 0 to 16 bar (100 °C), 0 to 10 bar (130 °C) |
| Process Connection | PG13,5 |
| Recommended Storage | Hamilton Storage Solution or 3 M KCl |
| Reference System | Hamilton Everef-L |
| Sample min Conductivity | 2 µS/cm |
| Sensitivity | 57 to 59 mV / pH at 25 °C |
| Sensor Family | Polilyte Plus |
| Sensor Output | Arc: Modbus, 4 to 20 mA |
| Serial Number | Yes |
| Steam Sterilizable | Yes, Maximum Temperature 130 °C |
| Temperature Sensor | NTC 22 kOhm |
| Wetted Parts | Glass \nFKM (Fluorocarbon Elastomer) |
| Zero Point | 0 ± 20 mV |
| Length | 325 mm |
| pH Glass | Hamilton Type H Glass |
| pH Glass Resistance | Moderate |
| Analog Interface 1 | 4 to 20 mA for pH, programmable |
| Analog Interface 1 and 2 | Galvanically not isolated; pulse width modulation 3.5 kHz |
| Analog Interface 2 | 4 to 20 mA for Temperature programmable |
| Baud Rate | 4800, 9600, 19200, 38400, 57600,115200 Bd |
| Configurable Values | pH: pH, mV; Temperature: °C, K, °F |
| Digital Interface | RS485 Modbus (Maximum 31 addresses) |
| Operating Voltage | 7 to 30 VDC Maximum 150 mW |