Hello , Sign In
Adrona Onsite+ is water polishing system which requires a connection to a water pre-treatment unit or a distilled water distribution system.It produces ultrapure water that complies with ISO 3696 Grade 1 water requirements.The system is recommended for laboratories with average daily consumption of water within 5-10 liters. As it contains portable 5 L storage tank, Onsite+ systems should be manually filled with pre-treated water.
| Product Type | Water Polishing System |
|---|---|
| Model | Onsite+ HPLC |
| Grade 1 Water Resistivity at 25 °C | 18.2 MΩ x cm |
| Grade 1 Water Conductivity At 25°C | 0.055 μS/cm |
| TOC Level | < 5 ppb~<50 ppb |
| Bacteria | < 0.01 CFU/mL |
| Endotoxins | < 0.15 EU/mL |
| Particles >0.22 μm | < 1/mL |
| Dispense Rate, Ultrapure Water | 1.5 to 2 L/min |
| Dimensions (WxDxH) | 30x44x64 cm |
| System Weight | 17 kg |
| Operating Weight | 22 kg |
| Feed Water | Pre Treated Water |
| Conductivity | <100 μS/cm |
| Pressure | 0.4 to 4.0 bar |
| Temperature | 5 to 35 °C |
| Display | Colour LCD |
| Dispenser | Dispense Port |
| Mains Voltage | 200 to 240 VAC |
| Power | 130 W |
| Fuse | 230 V, 3.15 A |
| Tank | Includes 5 L Portable Storage Tank |
| pH | 4 to 10 |
| Relative Humidity | 20 to 80% |
| Iron | <0.1 ppm as CaCO3 |
| Aluminium | <0.05 ppm as CaCO3 |
| Manganese | <0.05 ppm as CaCO3 |
| CO2 | <30 ppm |
| Fouling Index | <5 |
| Total Chlorine | <1 ppm |
| Langelier Saturation Index | <+0.2 |